free tracking

Air Temperature Control Unit SS-1200

The ATCU SS-1200 accommodates enclosure:
Area: Up to 2 m²
Volume: Up to 4 m³

Using solid state technology, the ATCU SS model accommodates enclosures up to 2 m² and will maintain the temperature at the Critical Point to 0.001 p-v (± 0.002) °C.

ACTU SS-1200 Specifications

Performance

Stability at Critical Point: 0.001 p-v
(± 0.002) °C.

Set-point: Ambient ± 1 °C.
Operating Conditions: 18 - 24 °C.
Air Flow Rates (m³/min): 10 - 25, adjustable.
Sound Level: 60 dBA.
Cooling Capacity: 1200 W.

Electrical 

Voltage (V): 115 V ± 10%.
Phase: 1.
Frequency (Hz): 50/60.
Max. Nominal Current - Cooling (A): 28 A.
Max. Nominal Power Consumption (kW): 2.6 kW.
Breaker/Fuse Size (A): 10.
Supply Power Wire Size (AWG): 12.

ENCLOSURE SIZE

Enclosure area:  Up to 2 m².
Enclosure volume:  Up to 4 m³.

PHYSICAL CHARACTERISTICS

Height: 1.6 m.
Depth: 0.8 m.
Width: 0.9 m.
Weight (kg): 140 kg.
HEPA Filtration: Optional

USER INTERFACE

Control:  Laptop PC.
Data storage:  Thermograph/PC.
Data recording format:  Excel, text. 
Channels:  Up to 8 channels of data.
Software:  Robust and user-friendly.
High/Low temperature alarm:  Yes.
Selectable sampling intervals:  Yes.
Moving average filtering:  Yes.
Adjustments:  Programmable set points.
Remote tuning from Præcis:  Yes.
User manuals:  English.